Watchdog
The AMI Watchdog provides a set of desirable features
that are needed for trace oxygen analysis at a remarkably low cost.
On the face of it, you might think that all you
really need is the oxygen sensor, an off-the-shelf cell block and a
single range amplifier so that you can plug it into your SCADA system or
flow computer. What could possibly go wrong?
As many people have discovered, there is more to it
than just that. A trace oxygen analyzer is an excellent leak detector.
The tiniest leak, due to improper assembly or incorrect choice of
sample system components, will render trace oxygen readings meaningless.
The sample system has to be able to provide a controlled flow both of
the measurement sample and a span gas and also allow the sensor to be
replaced rapidly (excessive exposure to air can damage or destroy a
trace oxygen sensor). The electronics has to be able to deal with the
wide range of possible oxygen readings from under 1ppm to 20.9% and
accurately compensate for temperature variations. The Watchdog provides
all of this – and extensive electronic capabilities - at a very low
cost.
Trace oxygen analyzers like the Watchdog are
essentially leak detectors. AMI’s patented cell block allows the
analyzer to be built with almost no possibility of internal leaking,
with minimal wetted volume and with front-panel sensor access. All
sample handling components – the flow meter, needle valve and a four-way
sample/span/off selection valve – are integrated into a solid metal
block. Connections between them are drilled passages. The result is a
highly reliable sample system with all necessary components provided,
and a very fast “come-down” time – when a sensor is replaced, or
calibrated on air with a 1 minute exposure (at 70°F), it takes about 15
minutes for it to come down to below 10ppm. Competitive analyzers
typically take hours.
The AMI oxygen sensor is inherently faster than
competitive sensors due to its metallic body and large active electrode
area. The standard T2 sensor is also naturally resistive to hydrogen
sulfide up to 10ppm H2S over its warrantied life, while the T4 is
capable of resisting up to 500ppm H2S, with no scrubber material
required, and hence no maintenance or leakage issues.
The sensor is immediately accessible on the front
panel of the analyzer, and can be replaced in seconds. It is not
necessary to expose the sensor to air unshorted while installing it as
is the case with most competitive analyzers (which significantly
increases their come-down time). As a result, although a span gas port
is provided, it is practical to accurately calibrate this analyzer on
air in less than a minute.
The electronic and software package provides a
standard set of features including ten analog output ranges. The output
may be field selected as either 4-20mA or 1-5V. The analyzer is
capable of operating off a battery/solar panel, and it logs power brown
outs and failures should they occur. It tracks the life of the sensor,
indicates the remaining life, logs the calibration history and logs the
oxygen readings over its operational (output) range, and also tracks
excursions of up to ten times that range.
The analyzer can be wall-mounted with 4 externally
accessible screws, or bolted to a 2” pipe using standard U bolts.

OVERVIEW
- Display reads oxygen from 0.05ppm up to 25.0% with no range changes
- Analog output can be configured to operate over any of ten ranges from 0-10ppm to 0-25% for best resolution
- Field selectable analog output - 1-5V or 4-20mA, isolated.
- Complete sample system built in to AMI’s patented cell block.
- Very rapid response time from air to low ppm levels.
- Easily replaceable sensor, no tools required.
- T-2 sensor resistant to 10ppm H2S standard. Optional T-4 resistant to 500ppm H2S.
- Simple, versatile installation.
- Operates off 10-28VDC with very low power consumption
- Analog output can be easily calibrated to a SCADA system or flow computer.
- USB connectivity to a PC: Allows complete access to the internal functions and settings.
- ModBus: Industry standard protocol over RS485.
- Oxygen Sensor life indication with calibration history.
FEATURES
- 10 user selectable output ranges to choose from.*
- High resolution 3 digit LCD.
- RFI protected.
- 1-5VDC and 4-20mA. isolated analog output signals.
- USB virtual comport and Modbus Bidirectional RS485 communication for advanced features.*
- Datalog – 10 days oxygen reading recording at 1 minute per sample*
- Calibration history – stores the last five calibrations with time, date, span factor and calibration gas.*
- Brown-out history – stores the last five brown-outs and recoveries.*
- Power up history – stores the last ten times the unit was powered up.*
- Advanced analog output calibration.*
(* Requires User Interface Software)
- Power requirements: 10-28VDC.
- Low minimum detection limit.
- Excellent repeatability.
- Fast upscale/downscale response times.
- Patented
Cellblock Technology: Integrates all components such as: flow control
valve, flow meter, 3-way calibration valve, Sample/Span/Off and
compression fittings to be an integral part of the cellblock,
eliminating tubing and fittings. The cell block also provides a compact
size, fast response time and front panel sensor access without the need
for tools.
- Area Classification: Designed to meet requirements for Class 1, Div. 2, Groups C,D application.
- Unaffected by changes in flow rate from 0.1 to 2.0 SCFH
- Wall mount or 2.0” pipe with standard pipe clamp.
- Compact size.
- 2 year warranty for analyzer, parts and labor.
- 6 month sensor warranty, life expectancy 1-2 years.