Model 2001LC
General purpose rack mount trace oxygen analyzer
The AMI model 2001LC is the ideal solution for
measuring trace oxygen in a wide range of applications in a general
purpose environment. In a small size, and at very low cost, it provides
a complete electronic package and sample system using AMI’s patented
cell block technology.
Unlike competitive analyzers, AMI sensor replacement
requires minimal downtime due to the front panel sensor access and the
patented AMI cell block. Typical time from a one minute air exposure to
10ppm is under 15 minutes.
- Display reads oxygen from 0.05ppm up to 25.0% with no range changes
- Analog output and alarms can be configured to operate over any of ten ranges from 0-10ppm to 0-25% for best resolution
- Analog output 4-20mA, isolated.
- Complete sample system built in to AMI’s patented cell block.
- Very rapid response time from air to low ppm levels.
- Easily replaceable sensor, no tools required.
- T-2 sensor resistant to 10ppm H2S standard.
- Simple, versatile installation.
- Operates off 10-28VDC with very low power consumption
- AC-DC converter provided
- 2 independent, fully adjustable alarm settings with relay contacts.
- Oxygen Sensor life indication.
- Trace oxygen – 0.00ppm to 25.0%
- Ultra-fast come-down sensor
- Positive pressure sample
- Patented cell block minimizes sample volume, fittings and tubing.
- Compact size – easy installation
- DC power 10-28VDC
- Low power
- Precision state-of-the-art electronics
- Isolated 4-20mA output
- 10 user-selectable analog output ranges
- Fully Programmable alarm logic (alarm delays, hold-off, etc.) with relay contacts
- Trace oxygen measurement from 0.05ppm to 25%.
- Unaffected by sample flow rate changes between 0.1 – 5 SCFH.
- Standard AMI T-2 trace electrochemical sensor.
- Panel mount.
- Compact size.
- 10 user selectable output ranges.
- 3 ½ digit LCD.
- 2 fully adjustable oxygen concentration alarms• RFI protected.
- Isolated 4-20mA. analog output signal.
- Power requirements: 10-28VDC <3 watts, supplied with a 115VAC to 12VDC adapter.
- Low original cost and virtually maintenance free over its entire life.
- Area Classification: Designed to meet General Purpose requirements.
- 2 year warranty for analyzer, parts and labor.
- 6 month sensor warranty.
- 2001LC Series Standard ranges:
- 0 – 10 ppm, 0 – 50ppm, 0 –100 ppm, 0 – 500 ppm, 0 – 1000ppm, 0 – 5000 ppm, 0 – 1%, 0 – 5%, 0 – 10%, 0 –25%
- Sensitivity: 0.5% of full scale
- Repeatability: +/- 1% of full scale at constant temperature
- Operating temperature: 41°F to 113°F
- Humidity: < 95%, non-condensing
- Maximum inlet pressure: 100psig.
- Operational conditions: Pollution degree 2, Installation category I I.
- Drift: +/- 1% of full scale in 4 weeks at constant temperature (dependent on sensor)
- Expected cell life: 9 months to 2 years.
- Response times:
- 90% of full scale in less than:
- 0 – 10 ppm 25 sec
- 0 – 100 ppm 10 sec
- 0 – 1000 ppm 10 sec
- Output: 4-20mA isolated.
- Alarm contacts: 230/117VAC @ 5A, or 28VDC @ 5A, resistive
- Power requirements: 24VDC <10W.
- Absolute Maximum Power voltage 28VDC
- Overall dimensions: 9” w x 5” h x 3” d
- Weight 5 lbs
|
|
|
|
|
|