Model 1000RS

Hazardous area portable trace oxygen analyzer
 
 
 
 
 
Trace oxygen analyzers are essentially leak detectors. Portable trace oxygen analyzers have the additional requirement that they should not be affected while being connected and disconnected to systems using tubing that - to begin with - is full of the very oxygen the analyzer is supposed to measure. If a trace oxygen sensor is exposed to air for any length of time, it will take a long time to recover back to a low reading, and if it is exposed to air repeatedly, its life will be affected and its response will get progressively worse. 
 
 
 
 
 
 
 
 
 
 
 
 
Model 1000RS
Model 1000RS
 
 
OVERVIEW
 
  • Trace oxygen measurement from 0.05ppm to 25.0%
  • Rugged, small, portable unit
  • Simple operation
  • Complete flow control and indication provided
  • Single quick-disconnect sample connection
  • Patented cell block provides purge capability
  • Very fast come-down time
  • Can be accurately calibrated on air
  • Sensor sealed off when not in use
  • Logs oxygen, temperature and battery voltage readings for subsequent recovery
  • Up to 350 hour battery life
  • Quickly and easily replaceable sensor, no tools required
  • Calibration history
  • Sensor life indication
  • Designed to be intrinsically safe
  • Approved by CSA International to UL and CSA standards as meeting requirements for Class 1, Div. 1, Groups B,C,D with a flammable gas sample.
 
DETAILS
 
  • Trace oxygen - 0.00ppm to 25.0%
  • Ultra-fast come-down H2S resistant sensor - up to 10ppm H2S resistant standard, 500ppm H2S resistant optional
  • Portable, battery powered
  • Patented cell block minimizes sample volume, fittings and eliminates tubing.
  • Integrated 4-way seal-off/bypass valve protects sensor from air contamination and purges system prior to measurement
  • Fastest come-down (time to measurement) in the industry
  • Precision state-of-the-art electronics
  • Data logging of oxygen readings, temperature and power supply voltage
  • 10 user-selectable analog output ranges
  • Calibration history
  • Calibration factor indicates remaining sensor life
 
 
 
 FEATURES
 
 
  • Patented Cellblock Technology: Integrates the sensor, needle valve, flow meter, and 4-way Sample/Bypass/On/Off valve, eliminating tubing and fittings.
  • Compact size
  • Rugged design
  • Front panel sensor access without the need for tools.
  • Rechargeable battery powered with up to 350 hours battery life
  • Fast upscale/downscale response times
  • Low minimum detection limit
  • Excellent repeatability
  • 3 ½ digit LCD
  • Auto-ranging display from below 1ppm to 25.0%
  • RFI protected
  • 0-1VDC analog output
  • Power requirements: 115/230VAC to 12VDC battery charger
  • Approved by CSA International to UL and CSA standards as meeting requirements for Class 1, Div. 1, Groups B,C,D with a flammable gas sample.
  • Unaffected by changes in flow rate from 0.1 to 2.0 SCFH
  • 2 year warranty for analyzer parts and labor
  • 6 month sensor warranty, sensor life expectancy 1-2 years
  • High level H2S resistant sensor optional
  • USB communication for advanced features*
  • Data logger with up to 10 days of 1 minute interval capacity*
  • 10 user selectable datalog ranges*
  • Datalog captures short term peaks of up to ten times the range selected*
  • Logs calibration history*
  • Logs power status*
  • Logs sensor oxygen exposure*
  • Sensor life indication*
  • * Requires AMI User Interface Software
 
 
 
 SPECIFICATIONS
 
 
  • 10 user selectable datalog and output ranges to choose from: 0-10ppm, 0-50ppm, 0-100ppm, 0-500ppm, 0-1000ppm,0-.500%, 0-1%, 0-5%, 0-10% and 25%
  • Digital display: 3 ½ digit LCD. Reads full scale from 0.00ppm to 25.0% independently of output range selection
  • Analog output signal: 0-1VDC
  • Represents the output range selected
  • Data logger: Logs data for 10 days @ 1 minute intervals, 20 days @ 2 minute intervals, etc. Represents the output range selected.
  • Power requirements: 115/230VAC to 12VDC battery charger. Rechargeable NiCad batteries, 350 hours continuous.
  • Minimum detection: 0.05ppm of oxygen
  • Repeatability: +/- 1% of range or +/- 0.2ppm of oxygen, whichever is greater
  • Operating temperature range: 25 to 115? F
  • Diurnal temperature specification: < +/- 3 % of scale over temperature range
  • 90% upscale response times: 10ppm – 25% <10 seconds 0-10ppm < 25 seconds
  • Typical downscale response: 1 minute exposure to air down to 10ppm: < 15 minutes
  • Approved by CSA International to UL and CSA standards as meeting requirements for Class 1, Div. 1, Groups B,C,D with a flammable gas sample.
  • Inlet gas pressure: 0.5 to 150psig
  • Gas connections: ¼” quick disconnect 316 S.S. compression fitting and ¼” nipple for exhaust
  • Wetted parts: 316 S.S. fittings, electroless nickel plated cellblock, gold plated contacts, acrylic flow meter and Viton O-rings
  • Unaffected by changes in flow rate from 0.1 to 2.0 SCFH
  • Portable dimensions: 4.5”W x 9.5”H x 4.5”D
  • Weight: 5 lbs
 
 
 
 
 